PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
October 2, 2025Circulation Arrhythmia and Electrophysiology4 citations

Modified Unipolar Return Pulsed Field Ablation in Ventricular Myocardium

View Full Paper
MTMaría TerricabrasPLPeter LombergarTETerenz Escartin

Key Points

  • The modified unipolar return PFA produced significantly deeper lesions compared to conventional methods, achieving depths of up to 11.6 mm.
  • Numerical modeling results demonstrated that the new PFA system had greater consistency in lesion depth and width across both healthy and infarcted myocardium.
  • Ablation in healthy swine resulted in lesion dimensions averaging 8.2 mm deep by 14.0 mm wide with 16 pulse trains targeting the left ventricle.
  • The study reveals potential for the modified PFA system in enhancing treatment outcomes in patients with myocardial infarction or other cardiac conditions.

Abstract

BACKGROUND: Various pulsed field ablation (PFA) parameters have been proposed to improve lesion depth. This study evaluated a modified unipolar return PFA system to create deep lesions in healthy and infarcted ventricular myocardia. METHODS: Numerical modeling was used to compare a modified unipolar return PFA system configuration with a conventional unipolar return (skin patch). We then performed ablation in 14 swine (5 with chronic myocardial infarction and 9 healthy). PFA lesions were created in the left ventricle using a focal catheter (4-mm tip) with a return electrode positioned in the inferior vena cava (biphasic, microsecond pulses of 1300 and 1500 V, 1–16 trains). Electroanatomical mapping guided ablation and lesion localization on magnetic resonance imaging were performed 48 hours post-ablation in the infarcted group and at 1 day, 7 days, and 6 weeks post-ablation in the healthy group. RESULTS: Numerical modeling demonstrated that the modified unipolar return PFA system produced deeper lesions with reduced variability compared with the skin patch. In healthy pigs (n=35 lesions), depths of 6.8±1.8 mm and widths of 11.5±4.7 mm were achieved with 8 pulse trains. Depths of 8.2±2.8 mm and widths of 14.0±4.7 mm were achieved with 16 trains. The maximum lesion depths were 8.8 and 11.6 mm for 8 and 16 trains, respectively. In the infarcted cohort (n=22 lesions), all lesions applied to scar tissue penetrated through fibrotic regions, with epicardial involvement observed in 57% of lesions. CONCLUSIONS: The modified unipolar return PFA system effectively creates large lesions and can achieve transmurality in healthy and infarcted animals. Compared with conventional unipolar, it may offer greater lesion depth, width, and consistency.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Terricabras et al. (2025) studied this question.

synapsesocial.com/papers/68de84c45b556a9128e1c05chttps://doi.org/10.1161/circep.125.014006
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Ablation of Myocardial Tissue With Nanosecond Pulsed Electric Fields2015 · 67 citations
  2. 2Feasibility, Efficacy, and Safety of Radiofrequency Ablation of Atrial Fibrillation Guided by Monitoring of the Initial Impedance Decrease as a Surrogate of Catheter Contact2015 · 49 citations
  3. 3Minimal coronary artery damage by myocardial electroporation ablation2012 · 115 citations
  4. 4Electroporation of the heart2005 · 98 citations
  5. 5In vivopulsed-field ablation in healthy vs. chronically infarcted ventricular myocardium: biophysical and histologic characterization2023 · 42 citations