Observational analysis compared smartwatch and CPET-measured VO2max in non-athletes, indicating limitations for clinical use.
Abstract Background. Maximal oxygen consumption (VO₂max) is the gold standard for assessing cardiorespiratory fitness (CRF), and it is a strong predictor of clinical outcomes, but it is measured during cardiopulmonary exercise testing (CPET), which is expensive and time-consuming. Smartwatches provide a great opportunity to assess CRF at scale. Although smartwatches have been previously validated in young and physically active individuals, their accuracy in sedentary individuals remains undetermined and this limits the use smartwatch VO₂max in large clinical or epidemiological studies. This study aims to determine the accuracy of smartwatch VO₂max in a large group of individuals with varying underlying CRF. Methods. N=203 adults (160 (79%) female; median (interquartile range; IQR) 55 (32,61) years old) from the CONVALESCENCE case-control study of long COVID, performed a maximal CPET on a semi-recumbent ergometer and wore a consumer-grade smartwatch for three months. Smartwatch-estimated VO₂max was compared to reference VO₂max measured by CPET and predicted VO₂max calculated using standard non-exercise equations. Agreement was assessed using Bland-Altman analysis (Bias [Limits of Agreement; LoA]), error (ΔVO₂max), absolute error (|ΔVO₂max|), and Pearson’s correlation coefficient (r). Results. CPET-measured VO₂max was 22.2 (17.5, 27.3) ml/kg/min. Smartwatch-estimated VO₂max correlated moderately with CPET-measured VO₂max (r=0.65) and strongly with VO₂max predicted using non-exercise equations (r=0.92). Bland-Altman plots showed a large positive bias and wide LoA (14.5 [2.2, 26.8 ml/kg/min]). Inaccuracy of smartwatch-estimated VO₂max (ΔVO₂max) linearly increased with increasing mismatch between expected and measured CRF (correlation with %pVO₂max r=-0.81). The absolute error was larger in those with lower-than-expected CRF (%pVO₂max<100%) and in those engaging in light intensity activities, while no differences were observed when comparing participants with or without long COVID or symptoms of fatigue. Multiple linear regression analysis of the absolute error identified the mismatch between expected and measured CRF (%pVO₂max) as the main contributor to smartwatch inaccuracy. Conclusions. In non-athletes, smartwatch-derived VO₂max correlated with reference values but did not provide additional accuracy to VO₂max prediction based on sex, age, weight and height. Inaccuracy significantly increased in those with lower-than-expected CRF. These findings highlight current limitations in the use of smartwatch VO₂max for clinical applications and large epidemiological studies. Trial registration number: Z6364106/2021/06/04.
No takes yet. Share an insight, caveat, or question.
Jamieson et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: