This assessment compares HbA1c levels in patients using DCA Vantage and Atellica DCA methods against the European Reference Laboratory, suggesting high correlation.
Background Measurement of HbA1c point-of-care testing (POCT) provides rapid test results during patient/doctor office visits. This added convenience also has potential for broader reach to low-access populations. The European Reference Laboratory for Glycohemoglobin (ERL) has seven different reference methods with different measurement principles which are calibrated with IFCC secondary reference material. Comparisons of the DCA Vantage and Atellica DCA HbA1c methods were analyzed as compared to the ERL value-assigned samples, which were designed to evaluate and standardize results across multiple POCT methods. Methods The DCA Vantage system HbA1c cartridges (CLIA-waived) and Atellica DCA HbA1c Dx (outside the United States) assays are based on a latex immunoagglutination inhibition methodology. To compare these methods a clinical specimen set was acquired from the ERL (N=40, value assigned by ERL). Method comparison studies were performed per CLSI EP09c 3rd edition, 2018. Two replicates of each of the 40 clinical samples were run on each system. All data were used for evaluation in this study. Value assigned samples were reviewed using both IFCC units (mmol/mol) and NGSP units (%), which were derived from the former. Statistical analysis of DCA Vantage vs ERL and Atellica DCA HbA1c vs ERL was performed using a linear mixed effects “means as outcomes” model and coefficient of determination (pseudo-R2). Analysis of the DCA Vantage vs Atellica DCA HbA1c methods was performed using weighted-Deming regression analysis. Results Comparison of the DCA Vantage to ERL methods of testing HbA1c level yielded y = 0.9726 x + 0.1766 % (y = 0.9726 x+ 1.2861 mmol/mol) at 4.62% to 10.69% (27.00 to 93.34 mmol/mol) with R2=0.9999. Also, comparing the Atellica DCA to ERL method of testing HbA1c levels yielded y = 0.9970 x + 0.0702 % (y = 0.9970 x+ 0.6995 mmol/mol) at 4.62% to 10.69% (27.00 to 93.34 mmol/mol) with R2=0.9983. Finally, comparison of the DCA Vantage to Atellica DCA methods of testing HbA1c levels, using weighted-Deming regression analysis, yielded y = 1.00 x + -0.01 % (y = 1.00 x+ 0.21 mmol/mol) at 4.65% to 10.76% (27.32 to 94.11 mmol/mol) with r=0.999. Medical decisions, based on American Diabetes Association, are made at 5.7%, marking the start of pre-diabetes, and up to 6.5% and higher, diagnosing Diabetes type 2 for most people. The 95% prediction ranges observed at these medical decision levels, as compared to the ERL method for the DCA Vantage and Atellica DCA, were determined: at 5.7% (5.40%,6.04% and 5.55%,5.96% respectively) and 6.5% (6.18%,6.81% and 6.35%,6.75%, respectively). Conclusion The Atellica DCA and DCA Vantage HbA1c assays strongly correlate to the European Reference Laboratory method and to each other.
No takes yet. Share an insight, caveat, or question.
Yost et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: