Why the study?
The analytical performance of the DCA Vantage point-of-care HbA1c instrument has been questioned, and little is known about the effects of changing reagent lot, instrument, and operator.
Does the Siemens DCA Vantage POCT instrument provide comparable analytical performance to routine HPLC methods for HbA1c measurement in a clinical setting?
Does the Siemens DCA Vantage POCT instrument provide comparable analytical performance to routine HPLC methods for HbA1c measurement in a clinical setting?
The DCA Vantage HbA1c POCT instrument showed poor analytical performance with significant lot and inter-device variation, emphasizing the need for rigorous validation and cautious clinical interpretation.
POCT HbA1c imprecision exceeds limits with lot and device effects; leaves open reliable clinical use pending rigorous validation.
Objectives The glycated haemoglobin fraction A 1c (HbA 1c ) is widely used in the management of diabetes mellitus, and the Siemens DCA Vantage™ point-of-care testing (POCT) instrument offers rapid HbA 1c results even far from a clinical laboratory. However, the analytical performance has been questioned, and not much is known about effects of changing reagent lot, instrument and operator. We therefore compared the analytical performance of the DCA Vantage™ with established routine methods (Tosoh G8/G11 ion exchange HPLC) in a true clinical setting at two Danish hospitals. Methods We extracted all routine clinical HbA 1c results incidentally drawn from the same patient within 48 h (n=960 pairs) and evaluated the effect of reagent lot, operator and instrument. We also performed a prospective method comparison in our diabetes out-patient clinic (n=97). Results The critical difference (CD) between two POCT results varied between 5.14 and 6.61 mmol/mol (0.47–0.55%), and the analytical imprecision of the DCA Vantage™ (CV A ) was >3%. Significant effect of reagent lot and inter-instrument differences were found, whereas no effect of operator was seen. Conclusions The DCA Vantage™ HbA 1c analysis does not fulfil the prevailing analytical performance specifications, but rigorous validation of new reagent lots and continuous recalibration of instruments may potentially improve the precision substantially. Our findings, therefore, clearly emphasise the necessity of a close collaboration between clinicians and laboratory professionals in the POCT field. Finally, POCT HbA 1c results should always be interpreted together with other measures of glycaemic control to avoid inappropriate change of patient treatments due to measurement uncertainty.
No takes yet. Share an insight, caveat, or question.
Abildgaard et al. (2021) studied this question.