Answers are grounded in this paper and cite their sources.
Patients with diabetes often experience blood glucose fluctuations, making monitoring crucial. Traditional blood sampling methods pose risks of infection and pain. An alternative non-invasive approach using urine tests has been explored. Recent studies highlight microfluidic paper-based analytical devices (µPADs) as convenient, simple, and easily fabricated tools for non-invasive glucose measurement. This study aims to develop a concept of measuring glucose in artificial urine using 3D-µPADs in a non-enzymatic manner by utilizing glucose as a reducing agent for silver nanoparticle (AgNPs) formation. Embedding three-dimensional connectors in µPADs links the sample and detection zones to limit reagent mixing and improve glucose detection resolution. The optimal conditions were NaOH 10 M, starch 1%, and AgNO3 30 mM, with sample and detection zone volumes of 10 and 9 µL, respectively. The fifth reaction sequence involved AgNO3 in the detection zone and a solution of glucose, NaOH, and starch in the sample zone at 1:1:1 volume ratio. The reagent drying time was 15 min, with immobilization once and reaction time of 9 min. The method showed excellent linearity (R2 = 0.9905), precision (%RSD = 4.27%), accuracy (77.32–92.58%), and limit of detection (11.11 mg/dL).
No takes yet. Share an insight, caveat, or question.
Fahmi et al. (2024) studied this question.