OBJECTIVE To assess the performance of transpulmonary thermodilution (TPTD) using room-temperature saline (CO RT ) and waveform-derived continuous CO (CCO) compared with TPTD using iced saline (CO ICED ) as the indicator for measurements of CO in isoflurane-anesthetized dogs. METHODS 8 Beagles aged 1 to 2 years (7.4 to 11.2 kg) were enrolled in this experimental study from March 21 to 31, 2023. Dogs were anesthetized with 0.01 mg/kg acepromazine, 5 to 6 mg/kg propofol, and isoflurane and were mechanically ventilated. Dogs were instrumented with a central venous catheter and a femoral arterial catheter equipped with a thermistor. The CO ICED , CO RT , and pulse wave-derived CCO values were obtained at baseline, during infusions of phenylephrine and norepinephrine, and during blood withdrawal and replacement. Data were analyzed with a mixed effect model, Bland-Altman plots, and concordance. Percent error was calculated. P < .05 was used for significance. RESULTS Data were collected from 8 dogs. Significant effects of time and the interaction of time and method were found. Bland-Altman plots showed negligible bias with limits of agreement between −0.35 and 0.25 L/min for CO RT versus CO ICED and −1.23 and 1.15 L/min for CCO versus CO ICED . Percent errors were 17.7% and 66.6%, respectively. In the 4-quadrant plots, the concordance rate was 95% and 68% for measurements obtained with CO RT and for CCO, respectively. CONCLUSIONS Transpulmonary thermodilution using room temperature saline was accurate and able to track changes in CO. Continuous CO had a large percent error and low tracking ability. CLINICAL RELEVANCE Transpulmonary thermodilution using room temperature saline is reliable for monitoring CO and obviates the need for iced preparations in clinical scenarios.
No takes yet. Share an insight, caveat, or question.
Cannarozzo et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: