Key result
Thermodilution cardiac output measurement during therapeutic hypothermia requires a temperature difference of at least 10°C between the patient's blood and the injected solution to ensure accuracy.
Why the study?
Does thermodilution cardiac output measurement provide accurate results in patients undergoing therapeutic hypothermia?
Does thermodilution cardiac output measurement provide accurate results in patients undergoing therapeutic hypothermia?
Thermodilution cardiac output measurements during therapeutic hypothermia may be inaccurate due to thermal noise, requiring careful evaluation of thermodilution curves to ensure a stable baseline.
Answers are grounded in this paper and cite their sources.
Urges caution interpreting thermodilution cardiac output during therapeutic hypothermia; leaves open validation of temperature-adjusted protocols in prospective studies.
Q When using the thermodilution method for cardiac output measurements, what temperature difference in degrees Celsius is required for accuracy? When we have a patient on the therapeutic hypothermia protocol and are using the thermodilution method for calculating cardiac output, the patient’s temperature may be as low as 32°C to 33°C with the injectate at room temperature (21°C–22°C). Is this enough of a difference?A Teresa Wavra, RN, MSN, CNS, CCRN, and Mary Kay Bader, RN, MSN, CCRN, CNRN, reply:Thermodilution cardiac output (TDCO) is based on the principles of dilution. The TDCO method uses a known quantity of cold solution to create a thermal deficit. A bolus of sterile solution that is colder than the patient’s blood is injected through the proximal port of the pulmonary artery catheter. The solution mixes with the patient’s blood, passes the tricuspid valve through the right ventricle and via the pulmonary valve into the pulmonary artery. The thermistor located 10 cm from the tip of the catheter senses the change in blood temperature. A curve that shows the change in temperature over time is calculated, and a computer converts this value into a measurement of cardiac output (Figure 1).For TDCO to be accurate, the signal-to-noise ratio must be adequate for the cardiac output monitor to sense a change in temperature over time. The signal is the temperature difference between the solution injected and the patient’s blood.1 The signal requires a temperature difference between the patient’s blood and the injected solution of at least 10°C and assumes that the temperature within the artery is stable during the measurement period.2 In clinical practice there may be temperature changes that are not produced by the solution injected into the patient’s blood. These temperature variation are considered thermal noise.Thermal noise, including hypothermia, influences the accuracy and reproducibly of TDCO measurements.3 Bazaral et al3 showed that alterations in pulmonary artery temperature caused substantial underestimation of cardiac output. Latson et al4 concluded that respiratory variations in the temperature of blood in the pulmonary artery are transiently increased in many patients after cardiopulmonary bypass and that this increased thermal noise may cause significant errors in TDCO measurements. The potential for errors can range from 15% to 50%.4Limitations of TDCO include the technical aspect of injecting the solution. Normally 10 mL of room-temperature solution is adequate, but cold solution is recommended in low-flow and hyperdynamic states.5 Variation in the volume of injectate and the speed of injection can produce differences between measurements. Rapid changes in temperature induced by rapid administration of fluid (>1 L/h), use of a warming blanket, and extra-corporeal oxygenation decrease the accuracy of cardiac output measurements. In tricuspid regurgitation, temperature is modified by regurgitation of blood into the right atrium and causes overestimation or underestimation of cardiac output.6Even if the circumstances are ideal, TDCO measurement has an error rate of 1%. Errors in measurement of cardiac output when the thermodilution method is used may be related to technical factors such as operator error, temperature transduction, baseline temperature variation in pulmonary artery blood, and instrument inaccuracies as well as the patient’s pathological conditions (eg, intracardiac shunt, abnormal hematocrit, and cardiac arrhythmia).5In contrast to TDCO, continuous cardiac output uses a 1-cm thermal filament with low levels of heat energy being transferred to the blood according to a pseudo-random binary (on/off ) sequence.5 The increase in temperature produced in the right ventricle is detected at a distal thermistor.7The heat signal is processed over time, and the thermodilution curve is rebuilt. The limitation of the thermodilution technique also applies to continuous cardiac output monitoring, but since the latter technique avoids interindividual variations in volume and speed of infusion of the bolus, its results show better reproducibility. Incorporation of the mean signal over a period of time could decrease the influence of the usual factors that limit the thermodilution technique.5,8During mild hypothermia, the changes in patients’ blood temperature may affect the signal-to-noise ratio and the accuracy of the measurement. The drift in the baseline temperature in the pulmonary artery is readily detected on examination of the thermodilution curves. To avoid using incorrect data, only thermodilution curves that begin and end on a stable baseline should be used (Figure 2).3Further investigation is needed to determine the validity and reliability of TCDO measurements in patients during the induction, maintenance, and rewarming phases of mild therapeutic hypothermia. Changes in hemodynamics, electrolyte levels, and intravascular fluid shifts occur during all phases of mild hypothermia. Recognizing these changes is of paramount importance in maintaining hemodynamic stability. A consistent method must be used to evaluate cardiac output/cardiac index so that the clinical hemodynamics of the patient during the 3 phases of temperature modulation can be regulated. The use of mild hypothermia after cardiac arrest is expanding in centers around the world. Nursing studies exploring the best methods for measuring cardiac output/cardiac index are needed to ensure that practitioners are using reliable data for monitoring the effects of hypothermia in critically ill patients.
No takes yet. Share an insight, caveat, or question.
Wavra et al. (2009) conducted a letter in Therapeutic hypothermia. Thermodilution cardiac output measurement was evaluated. Thermodilution cardiac output measurement during therapeutic hypothermia requires a temperature difference of at least 10°C between the patient's blood and the injected solution to ensure accuracy.