In this brief, a <tex-math notation="LaTeX">2ⁿᵈ </tex-math> -order <tex-math notation="LaTeX">gₘ - C </tex-math> lowpass filter with a practical input linear range of <tex-math notation="LaTeX">400~ mVP </tex-math> dedicated to ECG signal acquisition is proposed. This filter employs a degenerative source-coupled-pair circuit as a <tex-math notation="LaTeX">gₘ </tex-math> cell. It enhances the linear input range by a factor of <tex-math notation="LaTeX">× 4 </tex-math> compared with the source follower filter. Additionally, to mitigate the effect of current source mismatch, a dynamic element matching technique is applied. By doing so, HD2 is suppressed more than 1.5 dB over the entire passband frequency. This proposed filter is implemented in a <tex-math notation="LaTeX">0.18~μ </tex-math> m CMOS process. It offers a 250-Hz bandwidth with input-referred noise and a dynamic range of <tex-math notation="LaTeX">82.1~μ V_ rms </tex-math> and 67.34 dB, respectively. The power consumption of 38.4 nW is achieved with a 1.2 V supply. Compared with other recent nano-power filters, the proposed filter provides the highest linear input range with competitive Figure-of-Merit to the top-tier designs. It is therefore beneficial to the practical implementation of a low-power ECG acquisition system.