Case report reveals significant discrepancies in lactate levels due to β-hydroxybutyrate interference, indicating potential diagnostic risks.
Introduction Laboratory tests play a crucial role in diagnosing and managing illness. Accurate interpretation, which includes evaluating false-negative and false-positive results, is essential for guiding appropriate clinical interventions and ensuring patient safety. Methods We present a case involving clinically significant discrepancies in lactate measurements in a patient with acute myeloid leukemia and diabetic ketoacidosis. Results These discrepancies were caused by interference from elevated levels of ketone bodies—specifically, β-hydroxybutyrate—which may have delayed recognition of the severity of the patient’s condition. Discussion Dilution and spike experiments suggested potential interferences in core laboratory instruments using the Trinder reaction, leading to inaccurate lactate readings.