Case report reveals spurious bicarbonate elevation in serum sample due to enzyme interference, highlighting clinical implications.
Analytical interferences in automated chemistry analyzers can produce misleading results with potential clinical implications.We report a serum sample showing falsely elevated bicarbonate value associated with markedly increased lactate dehydrogenase (LDH) activity on the Beckman Coulter DxC 700 AU analyzer.Serum bicarbonate exceeded 45 mmol/L with a dilution flag, inconsistent with other analytical methods.The spurious elevation was most probably due to LDH interference affecting the NADH-dependent enzymatic reaction used in the bicarbonate assay.Awareness of this interference pattern is essential to prevent misinterpretation and unnecessary clinical concern.
No takes yet. Share an insight, caveat, or question.