Key points are not available for this paper at this time.
CaliberMRI’s multiparametric quantitative breast MRI phantom was scanned with a standard protocol, including DWI, at seven clinical sites as part of a quality control program. Protocol adherence was assessed. ADC values for fibroglandular and malignant breast tissue DWI mimics were calculated and corrected for temperature using the qCal software, and summary reports were generated. Inter- and intra-site variations in ADC derived from DWI phantom scans acquired longitudinally over the last 16 months provided information that can be used to troubleshoot imaging acquisition issues with sites and vendors as part of a phantom-based QC program.
Wilmes et al. (Wed,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: