Liver fat and R2* quantification has been extensively validated at 1.5T and 3T. Recently, there is renewed interest in lower-field MRI because of potential advantages such as a larger bore and lower costs. However, it is challenging to acquire images with sufficient signal-to-noise ratio for accurate fat and R2* quantification at 0.55T. Previous random matrix theory (RMT)-based approaches leverage Gaussian noise characteristics to markedly reduce the noise without compromising the tissue signal. In this work, we investigated a multi-coil multi-contrast RMT-based denoising approach which is compatible with parallel imaging and we demonstrated improved liver fat and R2* quantification at 0.55T.
No takes yet. Share an insight, caveat, or question.
Shih et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: