Answers are grounded in this paper and cite their sources.
We showed a new approach of MR temperature mapping without the oil-based phase reference of the conventional proton resonance frequency shift method (cPRFS), named as delta field-corrected temperature mapping (DFCTM). The performance of the DFCTM was compared with measurements using fiber optic thermal sensor and the cPRFS at the human-tissue mimicking agar-gel phantom. The sensor–measured temperature range was 6.05 ℃, while the DFCTM exhibited about 5.93 ℃. It exhibited a difference of -0.12 ℃ and the RMSE of the DFCTM was 0.16 ℃ for the entire experiments, which means that it could realize a temperature monitoring without oil phantoms.