This study describes a new method to improve RF safety of implantable medical devices located outside of the imaging region by using a secondary resonator (SR) to reduce electric fields and corresponding specific absorption rate (SAR) during MRI. The SR is designed to produce opposing electro-magnetic fields (EM-fields) compared to the EM-fields made by a body coil at 3.0T. This study was performed using numerical simulations with ASTM phantom and human models, and corresponding experimental verifications with the ASTM phantom.
No takes yet. Share an insight, caveat, or question.
Park et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: