We show the feasibility of compressed sensing for MRSI using concentric ring trajectories. Four volunteers (three for 2D- and one for 3D-MRSI) were measured in about 9 minutes (2D-MRSI), and 14 minutes (3D-MRSI) using an FID-based MRSI sequence at 7 T. An iterative compressed sensing reconstruction including coil sensitivities and a total general variation spatial regularization was performed with retrospective undersampling factors of 1.5, 2.0 and 2.9. RMSE values were below 22 % and SSIM values above 0.8 for medium accelerations below 2.9 in the 2D case. The metabolic maps and spectra are similar to the gold standard without acceleration.