Proton therapy is preferred for its dose conformality to spare normal tissues and organs-at-risk (OAR) via Bragg peaks with negligible exit dose. However, proton dose conformality can be further optimized: (1) the spot placement is based on the structured (e.g., Cartesian) grid, which may not offer conformal shaping to complex tumor targets; (2) the spot sampling pattern is uniform, which may be insufficient at the tumor boundary to provide the sharp dose falloff, and at the same time may be redundant at the tumor interior to provide the uniform dose coverage, for example, due to multiple Coulomb scattering (MCS); and (3) the lateral spot penumbra increases with respect to the depth due to MCS, which blurs the lateral dose falloff. On the other hand, while (1) the deliverable spots are subject to the minimum-monitor-unit (MMU) constraint, and (2) the dose rate is proportional to the MMU threshold, the current spot sampling method is sensitive to the MMU threshold and can fail to provide satisfactory plan quality for a large MMU threshold (i.e., high-dose-rate delivery).