Answers are grounded in this paper and cite their sources.
Multi-channel neural electrodes as a crucial means are of great significance for information exchange between the brain and computers. Herein, we present a 16-channel Si-based active neural probe system that achieves a monolithic integration between the electrodes and circuits in a single probe, making it a standalone integrated electrophysiology recording system. The ASIC prepared on a base (2×2mm2) is a 16-channel analog frontend (AFE) for neural recording, and each channel has a low-noise amplifier (LNA), a bandpass filter (BPF), a buffer and a current bias circuit. The 258 neural signal recording electrodes (22×24μm2) are densely packed on a 50 μm thick, 100 μm wide, and 3 mm long shank. The ASIC of neural probe, internal interconnecting wires are all implemented in commercial SMIC 0.18-μm CMOS technology. The neural probe system achieves a 3.6-μVrms input-referred noise (IRN) in a bandwidth of 1.1Hz-10kHz, 70.8-μW power consumption, 0.0785 mm2 area per channel, as well as an AFE gain of 58.1 dB. Furthermore, the impedances of the Au electrodes can be obtained as 0.5-2.1 MΩ at a frequency of 1 kHz. The functionality of a 16-channel silicon-based neural probe is validated in an in-vivo experiment on lab rats.
No takes yet. Share an insight, caveat, or question.
Gou et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: