Hemodialysis (HD) is a treatment supporting decreased kidney function, via a catheter inserted into the heart’s Right Atrium (RA). Recirculation is a source of inefficiency for treatment, where blood is dialyzed again due to poor catheter design. Lab-testing is still relatively unexplored, hence, a mechanical testing system was designed with the intention of providing a consistent and repeatable environment for testing HD catheters. System geometry was composed using a Computer-Aided Design (CAD) model of a heart, with the RA scaled to appropriate dimensions, and a PolyDiMethylSiloxane (PDMS) model produced through 3D printing and negative wax casting. Pulsatile blood flow was mimicked by peristaltic pumps driving a blood analogue (BA). Recirculation was induced by adding dyed BA to the system via the catheter and measured using a colorimeter. The developed platform was initially evaluated using two catheters, demonstrating the capability to accurately replicate atrial hemodynamic conditions. Two step-tipped catheters, A and B, were tested at 350 ml/min, producing recirculation values of 13.11% and 18.58%, respectively. The results exhibit the ability of the system developed to evaluate HD catheter performance, with the potential to explore a wider range of tip geometries relevant to clinical preference. Furthermore, this advancement towards an anatomically accurate lab-based test system could be paired with computational methods to progress the evaluation of such medical devices and enhance their development.