Key result
In an in vitro right atrial model, a multiorificed catheter aspirated up to 80% of incoming air when optimally positioned, compared to a maximal yield of 45-50% with a single-orificed catheter.
Why the study?
Does a multiorificed catheter improve air aspiration efficiency compared to a single-orificed catheter in an in vitro right atrial model?
Does a multiorificed catheter improve air aspiration efficiency compared to a single-orificed catheter in an in vitro right atrial model?
A multiorificed catheter positioned just below the SVC-atrial junction optimizes air aspiration in an in vitro right atrial model, suggesting a need to reappraise catheter design and placement.
Answers are grounded in this paper and cite their sources.
In vitro multiorificed catheter data are hypothesis-generating; clinical trials needed before any practice consideration.
A flexible Silastic casting of the human right atrium was developed to correspond to some in vivo human right atrium hemodynamic characteristics including chamber pressures, pulsatility, fluid output, and flow velocity. Using an infusion pump, air was introduced (10 ml in 30 s) into the superior vena cava of the model and aspirated via a catheter from different positions within the model atrial chamber. The tests were carried out at atrial inclinations of 60 degrees, 80 degrees, and 90 degrees from the horizontal and compared the aspiration efficiency of a single-orificed 16-gauge catheter to a 16-gauge multiorified (5 aperatures) catheter. Optimal air aspiration occurred with the multiorificed catheter tip positioned within the area 2.0 cm below the junction of the superior vena cava (SVC) and the atrial chamber at an inclination of 80 degrees. As much as 80 per cent of the incoming air could be aspirated under these conditions. At its optimal position the single-orificed catheter gave a maximal yield of 45 to 50 per cent aspiration when the tip was positioned 3.0 cm above the SVC and atrial chamber junction. Aspiration of air from mid right atrium (4.5 cm below the SVC-atrial junction) was poor regardless of the type of catheter used or atrial inclination. These data suggest a need for reappraisal of catheter design and placement.
No takes yet. Share an insight, caveat, or question.
Bunegin et al. (1981) studied this question. 16-gauge multiorificed catheter vs. 16-gauge single-orificed catheter was evaluated on Air aspiration efficiency. In an in vitro right atrial model, a multiorificed catheter aspirated up to 80% of incoming air when optimally positioned, compared to a maximal yield of 45-50% with a single-orificed catheter.