We are developing an imaging polarimeter by combining a fine-pixel CMOS image sensor with a coded aperture mask as part of the cipher project, aiming to achieve X-ray polarimetry in the energy range of $10$${x2013}30~keV. A successful proof-of-concept experiment was conducted using a fine-pixel CMOS sensor with a 2.5~μ m pixel size. In this study, we conducted beam experiments to assess the modulation factor (MF) of the CMOS sensor with a 1.5~μ m pixel size manufactured by Canon and to determine if there was any improvement in the MF. The measured MF was 8.32% ± 0.34% at 10~keV and 16.10% ± 0.68% at 22~keV, exceeding those of the 2.5~μ m sensor in the $6$${x2013}22~keV range. We also evaluated the quantum efficiency of the sensor, inferring a detection layer thickness of 2.67 ± 0.48~ μ m. To develop a more sensitive polarimeter, a sensor with a thicker detection layer, smaller pixel size, and reduced thermal diffusion effect is desirable.
No takes yet. Share an insight, caveat, or question.
Iwata et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: