Answers are grounded in this paper and cite their sources.
First discovered by Ernest Abbe in 1873, the resolution limit of a far-field microscope is considered determined by the numerical aperture and wavelength of light, approximately λ 2NA λ 2 N A . With the advent of modern fluorescence microscopy and nanoscopy methods over the last century, this definition is insufficient to fully describe a microscope’s resolving power. To determine the practical resolution limit of a fluorescence microscope, photon noise remains one essential factor yet to be incorporated in a statistics-based theoretical framework. We proposed an information density measure quantifying the theoretical resolving power of a fluorescence microscope in the condition of finite photons. The developed approach not only allows us to quantify the practical resolution limit of various fluorescence and super-resolution microscopy modalities but also offers the potential to predict the achievable resolution of a microscopy design under different photon levels.
No takes yet. Share an insight, caveat, or question.
Li et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: