Answers are grounded in this paper and cite their sources.
We report on a comparison of the performance of the Applied Biosystems® SeqStudioTM Genetic Analyzer with the 3500 Genetic Analyzer when analysing trace DNA samples. Totally, 162 profiles generated from 'trace DNA' were analysed on both instruments with comparable number of alleles detected, relative fluorescence units (RFU) and likelihood ratio (LR) values. The SeqStudioTM was found to be suitable for the analysis of trace DNA samples.
No takes yet. Share an insight, caveat, or question.
Griffin et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: