Answers are grounded in this paper and cite their sources.
Introduction: Obstruction of the left pulmonary artery (LPA) or descending aorta requiring late intervention is a rare complication after transcatheter closure with Piccolo device, but a multidisciplinary approach is often needed to discover and manage the patient when it does occur. We report our center’s experience with the Piccolo device, including 3 cases of late LPA stenosis requiring surgical debridement and pulmonary artery plasty. Methods: Retrospective chart review of 60 patent ductus arteriosus closures using Amplatzer Piccolo device in premature infants. Results: 60 cases of device closure in premature infants. Transient LPA stenosis was noted in 8% of patients, LPA stenosis requiring surgical intervention in 5% of patients. LPA velocity of 2.7 m/s at any point following device placement by echocardiogram, and right pulmonary artery (RPA) to LPA ratio of 1.5 to 1 by 2D echo are predictive indicators of need for surgical intervention. Differential lung perfusion scans may also provide important prognostic and follow-up data after intervention. Conclusion: Transient LPA stenosis is expected after device closure of a patent ductus arteriosus in premature infants. A multidisciplinary approach and vigilant follow-up are required to discover late complications including LPA occlusion and manage the patient when it does occur.
No takes yet. Share an insight, caveat, or question.
Jesus C. Jaile (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: