Answers are grounded in this paper and cite their sources.
The graphite electrode was changed by applying cyclic voltammetry to electropolymerized curcumin. Then, the polycurcumin-modified graphite electrodes (GECU) were used to determine Cd 2+ concentration. The results showed that the GECU electrodes had higher electrocatalytic activity toward Cd 2+ than bare graphite electrodes. Furthermore, the effect of the scan rates showed that the electrochemical process controlling the redox behavior of Cd 2+ on GECU is a diffusion-controlled electrochemical process. At Cd 2+ ranging from 0.22 to 11.03 ppm, the relationship between oxidation peak current (Ip) and Cd2 + concentration follows the linear regression equation of Ip = 73.024[Cd 2+ ] + 722.02 (μA, R 2 = 0.9936). It means that under experimental conditions, an increase in Cd2+ from 0.22 to 11.03 ppm can increase the Ip value. On the other hand, At Cd 2+ ranging from 11.03 to 55.07 ppm, the relationship between Ip and Cd 2+ concentration follows the linear equation of Ip = 24.227[Cd 2+ ] + 1212.6 (μA, R 2 = 0.9920). The LOD and LOQ were determined at low Cd 2+ concentrations, namely 0.86 ppm, and 2.84 ppm, respectively. Furthermore, the suggested modification has been used successfully and with satisfactory recoveries to determine the presence of Cd 2+ in laboratory liquid wastewater.
No takes yet. Share an insight, caveat, or question.
Marwati et al. (2024) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: