[Objective]: To optimize the graphite furnace atomic absorption spectrometry, compare the effects of blank pork matrix and 1% nitric acid as solvent on the detection of lead in pork, and establish a method for quantitative detection of lead in pork. [Methods]: The graphite furnace heating program and matrix modifier were optimized, and the method was screened. The results of three-level standard addition test and quality control product test of these two solvents were compared. [Results]: Under the screening conditions, with blank pork matrix as solvent, the correlation coefficient of the standard curve was 0.9999, the recovery rate of the three levels was 91.9% ~ 97.9%, and the coefficient of variation was 4.9% ~ 13.1%. The detection limit was 0.0134 mg / Kg. The test results of quality control products were within the allowable range; the test results with 1% nitric acid as solvent did not meet the standard requirements. [Conclusion]: Under the screening conditions, the accurate detection of lead in pork can be achieved by graphite furnace atomic absorption spectrometry with blank pork matrix as solvent.