This study presents the development and characterization of a graphene-based sensor integrated into a microfluidic chip for real-time monitoring of cell growth and viability. The sensor fabrication involved the metabolization of graphene from graphite using a simple and cost-effective method. The sensor design, created using Solidworks, featured electrodes capable of detecting environmental changes through impedance sensing. A mold was created using a cutter plotter to overcome challenges in achieving the desired sensor shape, and the electrodes were printed on a polyester (PETE) membrane. The conductivity of the electrodes was optimized through annealing, considering the temperature limits of the membrane. Annealing at 150 °C for 40 minutes yielded electrodes with desired conductivity and maintained membrane integrity. Compatibility with cellular growth was confirmed through cell culture experiments. The scaled electrodes were integrated into a microfluidic chip, and their performance was evaluated using cyclic voltammetry and electrochemical impedance spectroscopy. The results demonstrated the successful functioning of the electrodes within the chip. The developed graphene-based sensor offers promising applications in cellular studies and biosensing through real-time monitoring of cell growth and viability was achieved by measuring impedance changes resulting from cell attachment.
No takes yet. Share an insight, caveat, or question.
Ouedraogo et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: