Objective: To assess image artifact when imaging a cochlear implant (CI) with a conventional 3T MRI machine compared with a very low-field (0.064T) MRI. Patients: None. Intervention: Diagnostic study. Main Outcome Measure: Image artifact size associated with the CI affixed to an MRI phantom at very low-field 0.064T MRI versus 3T MRI. Results: The longest diameter of the image artifact was 125 mm for the 3T MRI and 86 mm for the 0.064T MRI, representing 45% longer image artifact generated in the 3T MRI. The actual volume of the imaging phantom was 1371 cm 3 . The volume of the image artifact was measured as 379 cm 3 in the 3T MRI, representing a loss of 27.6% of the actual volume of the imaging phantom. The volume of image artifact was measured as 170 cm 3 in the 0.064T MRI, representing a loss of 12.4% of the phantom volume. Conclusions: 3T MRI had better image quality. This result was not surprising given that larger magnetic field strength is known to provide higher resolution. There was 15% less image artifact generated in the very low-field MRI machine compared with a conventional 3T device. And there was also subjectively increased distortion of the imaging phantom at 3T MRI compared with the 0.064T MRI. With minimized safety concerns and a much lower cost than conventional 3T machines, very low-field scanners may find expanded clinical uses. This preclinical study explores the potential utility of very low-field MRI in scanning CI recipients.