Quantifies MRI artifact burden in auditory implants, suggesting bone-conduction options may improve image quality.
Purpose MRI in patients with auditory implants is limited by implant-related signal void, geometric distortion, and intravoxel dephasing. Artifact extent is often described qualitatively. We aimed to quantify MRI artifact burden from a novel transcutaneous bone-conduction implant (SentioTM) and to compare it with cochlear implant controls under routine, label-compliant 1.5 T clinical conditions. Methods We retrospectively analyzed MRI examinations of patients with a SentioTM implant ( n = 7) and cochlear implant controls ( n = 5). All scans were performed at 1.5 T, the manufacturer-approved field strength for the SentioTM Ti implant, on the same scanner using routine transverse turbo spin-echo sequences (T2-weighted and T1-weighted; 5‑mm slice thickness). Artifacts were quantified slice-by-slice with a custom Python-based DICOM pipeline. The artifact-to-brain ratio (ABR, %) and maximum void area (cm 2 ) were computed; ABRmax (peak slice) served as patient-level summary. Groups were compared using the Mann-Whitney U test and Cliff’s delta. Results SentioTM implants showed smaller and more confined artifacts. Median ABRmax was 1.1% [IQR 0.2] with a median maximum void area of 1.7 cm 2 [IQR 0.4]. Cochlear implant controls showed larger artifacts (ABRmax 2.6% [IQR 1.0]; median maximum void area 3.8 cm 2 [IQR 2.0]; p = 0.007–0.01; δ = 0.81–0.89). Supratentorial diagnostic image quality outside the immediate implant region was preserved in all T2w/T1w SentioTM examinations and partially preserved in 3/5 cochlear implant examinations. Conclusion Quantitative artifact assessment using ABRmax is feasible in routine 1.5 T MRI. Under identical protocol conditions, the SentioTM implant produced a smaller and more localized MRI artifact burden than the cochlear implant controls examined here. Routine turbo spin-echo imaging remained diagnostically useful outside the implant region.
No takes yet. Share an insight, caveat, or question.
Straub et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: