To accurately obtain the extremely weak magnetic field information of the test piece, this article first introduces the basic principles of the extremely weak residual magnetic field measurement technology. Then, an ultra-high precision magnetic shielding system has been developed, which is made of high permeability materials and equipped with a demagnetization coil system, and it can regularly eliminate the inductance of the entire shielding system. The effective shielding inner diameter of the magnetic shielding system is 0.8 m, the length is 1.6 m, and the magnetic field resolution is 0.0015 nT. This system can provide strong support for the measurement of extremely weak magnetic fields in materials and biomedical magnetic research.