This is the authors' abstract. We don't add key points for this paper.
A new superconducting circuit has been used in a sensitive magnetic susceptibility apparatus to detect changes in magnetic susceptibility as small as 10−10 cgs/g in a 104 Oe field, and in a low field magnetometer with which magnetic fields as low as 5×10−7 Oe have been measured. A low field superconducting shield for testing sensitive magnetometers has been developed and used to provide stable magnetic fields of less than 10−6 Oe, maintained stably for as long as six days. The measured attenuation of externally applied axial field changes is shown to be a factor of about 31 per shield radius. The stability of the persistent field of a Nb-25% Zr superconducting solenoid, observed over an 8 h period, indicated an average drift in the persistent solenoid field of less than one part in 1010/sec. Also, the attenuation of various high field magnetic shields of Nb-25% Zr foil was measured.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Deaver et al. (1967) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: