Answers are grounded in this paper and cite their sources.
Power cavitation imaging (PCI) is an emerging strategy for monitoring blood–brain barrier (BBB) opening, enabling spatial discrimination of cavitation intensity through power Doppler-analogous processing of passive cavitation images. Single-element cavitation detectors, while beneficial in signal monitoring for cavitation controllers, provide limited spatial information, conferring the need for an image-guided approach to ensure accurate cavitation localization and regulation. This study aims to evaluate the capability of PCI to spatially correlate real-time acoustic cavitation emissions with mechanical bioeffects as a predictor of P-selectin-targeted nanocarrier delivery. Preliminary sonoporation experiments were performed in vitro with brain tissue derived mouse endothelial cells and 3 kDa tetramethylrhodamine (TMR)-dextran. A focused ultrasound transducer (0.5 MHz, 5000 cycles/pulse, 5 Hz PRF, 30 second treatment duration) was used to sonicate cells in small volume (25 μl) cell suspension. Homogenous distribution of TMR-dextran in the cytosol and nucleus was observed via fluorescence microscopy at lower pressures (7% ± 2% fluorescent cells), while higher pressures had no significant differences above control (3% ± 2% vs 1% nontreated) due to microbubble destruction. Future experiments will correlate sonoporation with P-selectin expression and validate a linear array PCI system against contrast-enhanced MRI as a guide for P-selectin-targeted nanocarrier delivery in mouse models of medulloblastoma.
No takes yet. Share an insight, caveat, or question.
Onyemeh et al. (2024) studied this question.