Answers are grounded in this paper and cite their sources.
An analysis method was developed to allow for quantitation of muconic acid isomers (cis, cis and cis, trans) and aromatic compounds by ultra high pressure liquid chromatography paired with diode array detection (UHPLC-DAD). This was achieved by reproducible preparation of c,c and c,t isomers of muconic acid and chromatographic separation using a mobile phase gradient to separate various aromatic analytes and muconic isomers on a UHPLC reverse phase analytical column. In version 3 of this protocol a different C18 column option is used with the same gradient and mobile phases. This change was made due to inconsistancies in product specifications and QC with the previous column choice. Additional aromatic analytes were also added to the analytes quantified.