October 16, 2025ChemijaOpen Access

Beyond potentiometry: an amperometric approach to urea detection using antimony electrodes

View Full Paper
Ask AI
Bookmark
Share

Authors

JMJustas MiškinisInstitute of PsychologyMDMarius DagysVilnius UniversityJRJulija RazumienėVilnius University

Discussion

Loading...

Member takes

Overview

This amperometric biosensor shows strong correlation with a colorimetric assay in biological samples, suggesting effective urea detection.

Key Points

  • The biosensor demonstrated high sensitivity, with a range from 306.6 to 77.5 nA/mM depending on buffer conditions.
  • Stability tests revealed the sensor retained 65% of its initial activity after 10 days at room temperature.
  • Validation with biological samples indicated lower urea concentrations compared to a colorimetric method, affecting performance.
  • The developed biosensor is a cost-effective platform for urea detection, with potential in point-of-care diagnostics.

Cite This Study

Miškinis et al. (2025) studied this question.

synapsesocial.com/papers/68f10ecee6a12fd04289982fhttps://doi.org/10.6001/chemija.2025.36.3.1
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Highly Stable Potentiometric (Bio)Sensor for Urea and Urease Activity Determination2021 · 14 citations
  2. 2<i>Canavalia ensiformis</i> urease immobilization for urea biosensing: Rapid urea detection dipstick2026
  3. 3An Ultrasensitive Electrochemical Enzymatic Urea Biosensor Based on Aniline/N-Butyl Acrylate Conducting Polymer-Modified Screen-Printed Electrode2024 · 1 citations
  4. 4Biosensing of Urea with a Functionalized Gold Electrode for Health and Food Monitoring2025 · 9 citations
  5. 5Advances in Electrochemical Urea Biosensors: Trends and Future Prospects2025 · 4 citations