Answers are grounded in this paper and cite their sources.
Demonstrates a colorimetric biosensor's effectiveness in detecting urea, suggesting promising clinical applications.
A low-cost and rapid colorimetric biosensor for urea estimation was developed using immobilized urease from Canavalia ensiformis on a DEAE-cellulose membrane. Urease hydrolyses urea to ammonia, enabling colorimetric detection through pH changes monitored using phenol red (PR). The enzyme was purified via ammonium sulphate precipitation and anion exchange chromatography, yielding a specific activity of 841 U mL−1 and a molecular weight of 91 kDa confirmed by SDS-PAGE and western blotting. The purified enzyme exhibited 94–96% purity with an activity of 4.2 U µg−1. The immobilized enzyme (25.2 U mL−1) was co-adsorbed with PR (0.6 mg mL−1) on DEAE-cellulose strips. The biosensor demonstrated a limit of detection of 0.5 mM and a linear range of 0.5–12.5 mM in artificial serum samples. All experiments were performed in triplicate (n = 3), and results are reported as mean ± SD. The biosensor showed recovery of 92–95% compared with a spectrophotometric reference method, with no statistically significant difference (p > 0.05). The sensor remained stable for up to six months at 4 °C. Overall, the biosensor demonstrates excellent performance and stability, indicating its potential as a promising tool for urea detection in clinical and research settings.
No takes yet. Share an insight, caveat, or question.
Hussain et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: