Answers are grounded in this paper and cite their sources.
Adaptive filter demonstrates improved frequency response for ECG baseline wander removal, indicating enhanced signal clarity.
Baseline wander removal is a classical problem in ECG signal processing. We present a cascade adaptive filter to remove the baseline wander in the ECG preserving the overlapped deterministic low frequency components of the ECG, such as ST segment components. This cascade adaptive filter works in two stages. The first stage is an adaptive notch filter at zero frequency. The second stage is an adaptive impulse correlated filter that estimates the ECG signal correlated with the QRS occurrence. In both stages the LMS algorithm is used with different gain constants μ <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">1</inf> and μ <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</inf> . We analyse the frequency response of the filter as a function of the μ <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">1</inf> and μ <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</inf> parameters, selecting those more appropriated for baseline removal. Finally, the performance of the filter is studied on an actual ECG affected by baseline drift.
No takes yet. Share an insight, caveat, or question.
Laguna et al. (1992) studied this question.