December 8, 2025International Journal of Laboratory HematologyOpen Access

Clinical Evaluation of a Novel Point‐of‐Care Hematology Analyzer for Complete Blood Count With Differential

View Full Paper
Ask AI
Bookmark
Share

Authors

RSRyan C. SheanSBSterling T Bennett

Discussion

Loading...

Member takes

Implication

Proof-of-concept evaluation of a novel hematology analyzer identifies systematic differences in absolute neutrophil counts in various clinical settings, suggesting further investigation is needed.

Key Points

  • HemoScreen showed generally acceptable agreement with the Sysmex XN for most parameters.
  • Neutrophil percentage was 9.0% higher on HemoScreen, indicating a systematic difference.
  • Analysis involved 199 samples across inpatient, outpatient, emergency, and infusion center settings.
  • Further investigation is warranted to assess clinical impact before broader implementation.

Cite This Study

Shean et al. (2025) studied this question.

synapsesocial.com/papers/69401f002d562116f28f9d10https://doi.org/10.1111/ijlh.70032
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Analytical Performance Evaluation of Three Point-of-Care CBC Analyzers for Management of Clozapine Therapy in Ambulatory Psychiatry Clinics2023 · 6 citations
  2. 2Evaluation of the QBC Star centrifugal three-part differential haematology system2013 · 17 citations
  3. 3Verification and quality control of routine hematology analyzers2016 · 156 citations
  4. 4Diagnosis from the Blood Smear2005 · 412 citations
  5. 5The Top 25 Laboratory Tests by Volume and Revenue in Five Different Countries2018 · 99 citations