Traceable evaluation method reveals reduced periodic nonlinear error in micrometer range measurements, suggesting enhanced metrological applications.
Periodic nonlinear error (PNE) is a key factor limiting the accuracy of displacement measurement instruments. For grating interferometers, which are widely used in high-precision displacement measurement, reliable characterization of PNE is essential. Conventional evaluation methods that rely on higher-precision instruments become unsuitable for ultra-high-accuracy systems. Although the self-evaluation method based on Lissajous figures are commonly used, their results inherently depend on the grating parameters and the interferometric signals themselves, leading to a lack of traceability and reduced credibility of the results. In this work, we propose a traceable self-evaluation method for assessing the PNE of a directly traceable grating interferometer (DTGI). The DTGI employs a chromium (Cr) atom lithography grating, whose pitch is directly traceable to the atomic transition frequency of Cr (7S3 → 7P4), as its displacement reference, thereby overcoming the traceability deficiencies of conventional self-evaluation methods. Numerical simulations confirm the validity of the proposed method, and experiments performed on a laboratory-built DTGI demonstrate 10-picometer-level PNE after Heydemann correction within micrometer range displacements. These results confirm the method’s suitability for metrological applications requiring stringent linearity performance in nanometric displacement measurement.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: