On-site CT-derived FFR demonstrates ~84% accuracy against invasive FFR for detecting significant CAD.
Why the study?
Does a novel on-site CT-derived FFR algorithm accurately predict invasive FFR in patients with suspected chronic coronary syndrome and ≥50% diameter stenosis on CCTA?
Observational (n=38)
Does a novel on-site CT-derived FFR algorithm accurately predict invasive FFR in patients with suspected chronic coronary syndrome and ≥50% diameter stenosis on CCTA?
Effect estimate: median paired difference 0.083 (95% CI 0.063-0.10)
Absolute Event Rate: 0.84% vs 0.92%
A novel on-site CT-derived FFR algorithm demonstrated 100% sensitivity and negative predictive value for detecting functionally significant coronary artery disease compared to invasive FFR.
Background Computed tomography (CT)-derived fractional flow reserve (FFR) improves the diagnostic accuracy of coronary CT angiography (CCTA) but requires off-site transfer of images and long processing times. A novel deep learning-based, on-site FFR algorithm offers rapid (8-minute) assessment with promising diagnostic performance but requires external validation before clinical implementation. Purpose To examine the agreement and diagnostic performance of CT-derived FFR with invasive FFR (iFFR) as the reference for functionally significant coronary artery disease and to assess their correlation and diagnostic performance against myocardial perfusion with [15O]H2O positron emission tomography (PET) Methods Post-hoc analysis of the Danish Study of Non-Invasive Testing in Coronary Artery Disease 3 (Dan-NICAD 3) study. Patients with suspected chronic coronary syndrome and ³50% diameter stenosis on CCTA underwent [15O]H2O PET and invasive coronary angiography with three-vessel iFFR. The primary outcome was the agreement between CT-derived FFR and iFFR , and secondary outcomes included the diagnostic accuracy of CT-derived FFR in predicting abnormal iFFR and the diagnostic accuracy of CT-derived FFR and iFFR in predicting core lab-adjudicated ischemia. Results In total, 38 patients (mean age 64 ± 8.0 years, 76% male) with 87 vessels were included for analysis. iFFR ≤0.8 was present in 21/38 (55%) patients and 23/87 (26%) vessels. Median invasive and CT-derived FFR were 0.92 (IQR 0.80–0.97) and 0.84 (IQR 0.69–0-91) (median paired difference 0.083, 95% CI 0.063–0.10), respectively, with a rho of 0.74. Against invasive FFR ≤0.80 on a per-patient level, CT-derived FFR ≤0.8 had an accuracy of 84% (95% confidence interval (CI) 70–93%), sensitivity of 100% (95% CI 86–100%), specificity of 65% (95% CI 41–83%), positive predictive value (PPV) of 78% (95% CI 59–89%), negative predictive value (NPV) of 100% (95% CI 74–100%), negative likelihood ratio (LR) (LR-) of 0.0 (95% CI 0.0–0.21), and positive LR (LR+) of 2.8 (95% CI 1.5–5.4). In comparison with ≥50% diameter stenosis on CCTA, CT-derived FFR ≤0.8 against the reference of iFFR ≤0.8 had similar sensitivity, NPV, and LR- but higher diagnostic accuracy, specificity, PPV, and LR+ (Table 1 + Figure 1). Both CT-derived FFR and iFFR correlated weakly with stress myocardial blood flow (rho 0.30 and 0.36, respectively) and myocardial flow reserve (rho 0.20 and 0.38, respectively). The diagnostic accuracy of CT-derived and iFFR in predicting ischemia was 46% (95% CI 30–64%) and 61% (95% CI 42–76%), respectively, on a per-patient level, and 69% (95% CI 57–79%) and 75% (95% CI 63–84%), respectively, on a per-vessel level. Conclusions A novel on-site CT-derived FFR algorithm demonstrates fast and very high rule-out capability and incremental rule-in potential beyond diameter stenosis for functionally significant disease in patients with suspected stenosis at CCTA.
No takes yet. Share an insight, caveat, or question.
Soeby et al. (2026) conducted an observational in Suspected chronic coronary syndrome (n=38). On-site CT-derived fractional flow reserve vs. Invasive fractional flow reserve (iFFR) was evaluated on Agreement between CT-derived FFR and iFFR (median paired difference 0.083, 95% CI 0.063-0.10). An on-site CT-derived FFR algorithm demonstrated 84% diagnostic accuracy (95% CI 70-93%) and 100% sensitivity against invasive FFR for detecting functionally significant coronary artery disease.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: