Comparison of AC and DC amplifier effects on electroretinography recordings, indicating potential biases in data representation.
Almost all ERGs are obtained by using AC amplifiers, because these are more stable. However, the AC amplifier delivers a degraded first derivative of the DC signal at the cornea, resulting in a misrepresentation of both time and amplitude, as well as an entirely artifactual “negative overshoot” which does not exist at all. ERGs amplified by both AC and DC methods are presented for comparison. A new type of DC amplifier has permitted a new and greatly simplified geometric analysis of the ERG.
No takes yet. Share an insight, caveat, or question.
R. H. Peckham (1971) studied this question.