Key result
Modern speckle tracking software yields <1% absolute difference in GLS, enabling interchangeable patient follow-up.
Why the study?
Despite efforts by associations and industry, speckle tracking strain imaging has shown inter-vendor variation, prompting assessment of whether modern strain analysis yields improved reproducibility.
Do modern strain analysis software packages yield reproducible Global Longitudinal Strain (GLS) results across different vendors?
Do modern strain analysis software packages yield reproducible Global Longitudinal Strain (GLS) results across different vendors?
Modern commercially available strain software packages provide similar GLS measurements with <1% absolute difference, suggesting they may be interchangeable for serial patient follow-up.
Introduction Speckle tracking strain imaging has become a routine part of echocardiography. However, despite the efforts of multiple associations and industry, there has been variation in results between vendors. We aimed to determine if modern strain analysis yielded results with improved reproducibility. Methods Participants underwent echocardiography on three different ultrasound machines: GE E95, Philips CVx and Mindray M9. Global longitudinal strain (GLS) was calculated on each system and on vendor‐independent software TOMTEC. Absolute and relative differences in GLS were calculated. Results Of the 50 participants, GLS values were similar from each software (GE EchoPAC = −20.1% ± 2.0%, Mindray on‐cart Tissue Tracking QA‐LV = −19.7% ± 2.7%, Philips AutoStrain LV = −19.3% ± 2.8%, TOMTEC Autostrain [GE images] = −20.1% ± 2.6% and TOMTEC Autostrain [Mindray images] −19.8% ± 2.8%). All strain software analysis had moderate to excellent correlation. Furthermore, all GLS results had a < 1% absolute difference and < 10% relative difference between each other. Conclusion This study demonstrates that the performance of commercially available strain software is similar and may be interchanged for serial follow up of patients; however, this relies on having the most updated software version available.
No takes yet. Share an insight, caveat, or question.
Perry et al. (2026) studied this question. Modern speckle tracking strain software packages show similar GLS values with <1% absolute and <10% relative difference, enabling interchangeable use for patient follow-up.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: