We have previously investigated a hemodialysis training method in which magnetic particles are circulated in a dialysis circuit and occlusion is controlled by an external magnetic field to reproduce abnormal conditions. In this study, in order to expand the scope of application to various extracorporeal circulation training, we clarified the circulatory limit flow rate of magnetic particles in the ducts and the relationship between the duct occlusion rate and pressure. Furthermore, we investigated the circulation of magnetic particles and pressure changes in the CBP blood circuit and the ECMO blood circuit. As a result, the circulatory limit flow rate was 50 mL/min and 600 mL/min for duct inner diameters of 3 mm and 12 mm, respectively. Furthermore, the duct occlusion rate could be adjusted up to 100% for each duct diameter, and the pressure at that time could be increased to 400 mmHg, which is expected in clinical practice. In addition, since magnetic particles could circulate in the CBP or the ECMO blood circuit and occlusion could be controlled by the field at abnormal sites expected in clinical practice, the scope of application of the method can be expanded to various extracorporeal circulation training.