Observational analysis compares patient-specific designs with conventional plates, revealing superior stiffness and displacement performance.

Key Points

  • Both 3D reconstruction and generative design plates exhibited increased stiffness, outperforming conventional plates.
  • In bending tests, 3D reconstruction plates were 23.8% stiffer, while generative design plates increased stiffness by 10%.
  • Biomechanical tests demonstrated that patient-specific plates had 15% lower displacement than conventional plates under load.
  • Integration of poly(lactic acid) prototypes with metal additive manufacturing may enhance surgical decision-making.