Key result
Four tested metabolic carts show comparable intra-device reliability and inter-device consistency under standardized conditions.
Why the study?
Different metabolic carts used for cardiopulmonary exercise testing may introduce systematic variability, confounding longitudinal assessments and multicentre trials.
Are CPET parameters comparable across different metabolic carts and reliable within the same device in healthy adults?
Population
34 healthy adults (35 ± 11 years; 59% males)
Comparison
Intra-device reliability and inter-device comparability of four metabolic carts
Design
Prospective, single-centre, within-subject, repeated-measures, multi-device comparison study with randomized test order
Follow-up
Within 40 days (tests 2-10 days apart)
Authors
Loading...
Clinicians can rely on CPET data independent of cart chosen; confirms inter-device reliability for clinical and research use.
Are CPET parameters comparable across different metabolic carts and reliable within the same device in healthy adults?
Under standardized conditions, different metabolic carts yield comparable peak VO2 and VE/VCO2 slope (up to RCP) values, supporting their reliable use in longitudinal assessments and multicentre trials.
Willixhofer et al. (2026) studied Healthy adults (n=34). Cardiopulmonary exercise testing using four different metabolic carts vs. Intra- and inter-device comparison was evaluated on Intra-device reliability and inter-device comparability of CPET parameters (peak VO2, VE/VCO2 slope). Four different metabolic carts yielded comparable values for intra-device reliability and inter-device comparability of peak VO2 and VE/VCO2 slope up to the respiratory compensation point.