Key result
Four tested metabolic carts show comparable intra-device reliability and inter-device consistency under standardized conditions.
Why the study?
Different metabolic carts used for cardiopulmonary exercise testing may introduce systematic variability, confounding longitudinal assessments and multicentre trials.
Are CPET parameters comparable across different metabolic carts and reliable within the same device in healthy adults?
Are CPET parameters comparable across different metabolic carts and reliable within the same device in healthy adults?
Under standardized conditions, different metabolic carts yield comparable peak VO2 and VE/VCO2 slope (up to RCP) values, supporting their reliable use in longitudinal assessments and multicentre trials.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Clinicians can rely on CPET data independent of cart chosen; confirms inter-device reliability for clinical and research use.
AIMS: Cardiopulmonary exercise testing (CPET) is the gold standard for assessing exertional dyspnoea, providing peak oxygen uptake (VO2), and the VE/VCO2 slope. Different metabolic carts, may introduce systematic variability, confounding longitudinal assessments and multicentre trials.We evaluated intra-device reliability and inter-device comparability of CPET parameters. METHODS AND RESULTS: In this prospective, single-centre, within-subject, repeated-measures, multi-device comparison study 34 healthy adults (35 ± 11 years; 59% males) completed eight CPETs: two per cart, on four systems (MGC Ultima™ CPX, COSMED Quark CPET™, Schiller PowerCube® Ergo, Vyaire Vyntus™ CPX). Tests used randomized order, ramp protocols on cycle ergometer, and standardized calibration, masks, and timing (2-10 days apart; within 40 days). Breath by breath data were 10-second averaged. Analyses included paired tests for intra-device reliability, repeated-measures ANOVA for inter-device comparisons, and iso-workload evaluation at 25-100% of lowest peak workload (ISO-Wpeak). Intra-device reliability was high, with no significant differences in VO2, heart rate or workload at anaerobic threshold (AT), respiratory compensation point (RCP) or peak. Minor variances occurred in resting VO2 (COSMED: 56 mL/min, P = 0.0427) and peak ventilation (Medical Graphic 9 L/min, P < 0.05). Inter-device comparability showed no differences in peak VO2 and the VE/VCO2 slope calculated up to the RCP. The VE/VCO2 slope assessed as the full slope however, showed significant differences between machines (P = 0.0048). Across standardized ISO-Wpeak (25%, 50%, 75%, and 100% of ISO-Wpeak), VO2 values were largely comparable between devices, with significant differences only for VO2 (mL/kg/min) at 25% ISO-Wpeak (P = 0.022). CONCLUSION: Under standardized conditions, the tested metabolic carts yield comparable values both for intra-device reliability and inter-device comparability.
Willixhofer et al. (2026) studied Healthy adults (n=34). Cardiopulmonary exercise testing using four different metabolic carts vs. Intra- and inter-device comparison was evaluated on Intra-device reliability and inter-device comparability of CPET parameters (peak VO2, VE/VCO2 slope). Four different metabolic carts yielded comparable values for intra-device reliability and inter-device comparability of exercise capacity and VE/VCO2 slope up to the respiratory compensation point.