Key result
Jaeger Oxycon Pro with antibacterial filter shows no difference in VO2max versus COSMED Quark RMR.
Why the study?
Differences among metabolic carts during CPET can arise from internal algorithms, sensor technology, antibacterial filters, or obsolescence, prompting comparison between the Jaeger Oxycon Pro and Quark RMR.
Does the Jaeger Oxycon Pro with an antibacterial filter provide comparable metabolic and ventilatory parameters to the COSMED Quark RMR during cardiopulmonary exercise testing?
RCT (n=42)
Randomized order
Does the Jaeger Oxycon Pro with an antibacterial filter provide comparable metabolic and ventilatory parameters to the COSMED Quark RMR during cardiopulmonary exercise testing?
Answers are grounded in this paper and cite their sources.
The Jaeger Oxycon Pro with an antibacterial filter and the COSMED Quark RMR provide highly correlated and agreeable measurements during cardiopulmonary exercise testing, suggesting they can be used interchangeably.
Several factors, such as internal algorithm, sensor technology, use of antibacterial filters or obsolescence, might explain the differences among computerised metabolic carts when measuring gas exchange parameters during cardiopulmonary exercise tests (CPET). Thus, the main aim of this study was to compare the Jaeger Oxycon Pro and the Quark RMR during CPET in the context of an aerosol-borne infectious disease. Forty-two participants (19-54 years old) performed two non-consecutive maximal graded exercise tests on a cycle-ergometer on two counterbalance and different days using the Oxycon Pro, with and antibacterial filter, and the Quark RMR in a randomized order. Differences between devices were tested using a generalized linear model adjusted by Bonferroni, and correlation and agreement was assessed using Pearson’s correlation coefficient (R), intraclass coefficient correlation (ICC), Lin’s concordance correlation coefficient (CCC) and Bland-Altman plots. No significant differences were found between devices in any of metabolic or ventilatory parameters for cardiorespiratory fitness assessment (VO2max: 3131.3±882.1 vs. 3189.8±894.8, p=0.071, VCO2max: 3436.3±936.0 vs. 3550.3±1043.2, p=0.071, for Quark RMR and Oxycon Pro, respectively). However, Bland-Altman plots showed a trivial tendency towards Oxycon Pro overestimation relative to Quark RMR as air flow volume increases. Both devices showed strong correlation and high level of agreement during maximal and submaximal exercise intensities (Pearson’s R: 0.974 & 0.977; ICC: 0.985 & 0.987; Lin’s CCC: 0.971 & 0.974, for VO2 and VCO2, respectively). Therefore, providing consistency to CPET data comparison between both devices, and insight into whether the use of these metabolic carts could be interchangeable or combined when a single device cannot be used for CPET assessment.
No takes yet. Share an insight, caveat, or question.
Muñoz-Muñoz et al. (2023) conducted an RCT in Cardiopulmonary exercise testing (n=42). Jaeger Oxycon Pro with antibacterial filter vs. COSMED Quark RMR was evaluated on VO2max (p=0.071). The Jaeger Oxycon Pro with an antibacterial filter showed no significant difference in VO2max compared to the COSMED Quark RMR, with strong correlation between devices.