Key result
CRT-P implantation using a right-sided delivery system and bipolar coronary sinus lead successfully reduced QRS duration from 202 to 142 ms, improved ejection fraction to 49%, and resolved exertional symptoms.
Why the study?
Cardiac resynchronization therapy pacemaker placement and coronary sinus cannulation are notably challenging in dextrocardia, requiring unconventional strategies.
Case Report (n=1)
No
In a patient with dextrocardia and situs inversus totalis, deploying a right-sided delivery system and a bipolar coronary sinus lead from a left venous access point successfully achieved cardiac resynchronization and improved clinical and echocardiographic parameters.
Answers are grounded in this paper and cite their sources.
May guide CRT planning in dextrocardia; hypothesis-generating and should not yet change practice.
Background Cardiac resynchronization therapy pacemaker (CRT-P) placement and coronary sinus cannulation (CSC) are notably challenging in dextrocardia (DXC). Case summary A 56 year old female with DXC, sick sinus syndrome (SSS), paroxysmal atrial fibrillation (pAF), and heart failure reduced ejection fraction (HFrEF) presented for symptoms with exertion. Electrocardiogram (ECG) showed prolonged QRS necessitating CRT-P and CSC with right sided multipolar catheter and bipolar lead. Post operatively there was improvement in QRS, EF, and symptoms. Discussion CRT-P and CSC techniques in levocardia are well established. DXC requires unconventional CRT-P strategies. A right sided multipolar catheter is valuable for CSC in DXC. This DXC case utilized a right sided electrophysiology catheter and bipolar coronary sinus (CS) lead. The use of these devices ensured that if the prior right ventricular (RV) ICD lead failed, a standard high S1 lead would be compatible with pacemaker headers. Take Home Message A right sided CSC system is useful for implanting CS leads with a left-sided device in DXC. While it comes at the expense of losing different pacing vectors, a bipolar CS lead is essential if there is a risk that the pre-existing bipolar RV lead will fail.
No takes yet. Share an insight, caveat, or question.
Khasnavis et al. (2026) conducted a case report in Dextrocardia with situs inversus totalis and pacing-induced heart failure (n=1). Cardiac resynchronization therapy pacemaker (CRT-P) utilizing a right-sided delivery system and bipolar coronary sinus lead was evaluated on Successful CRT-P implantation and improvement in QRS duration, ejection fraction, and clinical symptoms. CRT-P implantation using a right-sided delivery system and bipolar coronary sinus lead successfully reduced QRS duration from 202 to 142 ms, improved ejection fraction to 49%, and resolved exertional symptoms.
Synapse has enriched 10 related papers. Consider them for comparative context: