This work demonstrates glucose sensing using an electrochemical biosensor, suggesting effective electron transfer mechanisms in glucose detection.
Facilitating efficient electrical communication between glucose oxidase (GOx) and electrode surfaces is important for the development of low-potential reagent-free glucose biosensors, yet unequivocal direct electron transfer (DET) for native GOx remains controversial because the FAD redox center is deeply buried within the protein structure. In this work, a one-step electrochemical co-deposition strategy was used to fabricate an electrochemically reduced graphene oxide/L-lysine (ERGO–Lys) nanocomposite on glassy carbon. The resulting interface exhibited a pair of well-defined redox peaks centered at −0.460 V (vs Ag/AgCl) and glucose-responsive behavior under rigorously deoxygenated conditions. Given the continuing debate surrounding native GOx DET at carbon nanomaterials, these results are interpreted conservatively as DET-like (pseudo-DET) bioelectrochemistry rather than definitive proof of true DET. Under anaerobic conditions, the biosensor showed a linear range of 5 μM to 9.0 mM, a sensitivity of 35.8 μA·mM⁻¹·cm⁻², and a limit of detection of 1.8 μM. The sensor also retained 90% of its initial response after two weeks and showed low interference at −0.40 V.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: