Key result
Cast iron skillet can mitigate pVAD-induced electromagnetic interference disrupting ICD telemetry.
Why the study?
Percutaneous ventricular assist device induced electromagnetic interference can disrupt implantable cardioverter-defibrillator telemetry.
Does placing a cast iron skillet over the heart resolve pVAD-induced electromagnetic interference with ICD telemetry in a hemodynamically unstable patient?
Case Report (n=1)
Does placing a cast iron skillet over the heart resolve pVAD-induced electromagnetic interference with ICD telemetry in a hemodynamically unstable patient?
Answers are grounded in this paper and cite their sources.
A cast iron skillet can act as a pseudo-Faraday cage to successfully shield pVAD-induced electromagnetic interference and restore ICD telemetry in hemodynamically unstable patients who cannot tolerate pump weaning.
BACKGROUND: Percutaneous ventricular assist device (pVAD)-induced electromagnetic interference (EMI) can disrupt implantable cardioverter-defibrillator (ICD) telemetry. We describe a novel mitigation strategy using a cast iron skillet. CASE SUMMARY: An 81-year-old woman with an ICD underwent pVAD-assisted percutaneous coronary intervention for cardiogenic shock. Postprocedure interrogation for slow ventricular tachycardia failed due to significant EMI. Because hemodynamic instability precluded reducing pVAD flow to mitigate interference, a cast iron skillet was placed over the heart. This pseudo-Faraday cage successfully restored telemetry. DISCUSSION: Although the "skillet method" aids durable left ventricular assist device interrogation, prior reports suggest inefficacy with pVADs, favoring flow reduction. We report the first successful use of a cast iron skillet to shield Impella EMI, offering a vital strategy for unstable patients unable to tolerate pump weaning. TAKE-HOME MESSAGES: pVAD systems can generate EMI, disrupting ICD telemetry. When hemodynamic instability prevents lowering pVAD flow, a cast iron skillet can assist in shielding EMI and restoring communication.
No takes yet. Share an insight, caveat, or question.
Shaheriyar et al. (2026) conducted a case report in Cardiogenic shock with pVAD-induced electromagnetic interference (n=1). Cast iron skillet was evaluated on Restoration of ICD telemetry. Placement of a cast iron skillet over the heart successfully restored ICD telemetry by shielding pVAD-induced electromagnetic interference in a hemodynamically unstable patient.
Synapse has enriched 10 related papers. Consider them for comparative context: