Key result
Stratifying diverse MRI experiments demonstrates equal temperature spread in a radiofrequency power versus B1+rms comparison near implanted conductive devices.
Design
Editorial
Answers are grounded in this paper and cite their sources.
This letter discusses methodological concerns regarding RF shimming instructions and experimental parameter control in evaluating MRI-induced heating of implanted devices.
We1Response to Authors’ Reply to Comments received Oct. 8, 2025. Complete correspondence appears in the March 2026 issue. thank the authors [1] for their response. This final reply focuses on three key sentences: 1)The following incorrect statement: “there are no user-controlled RF shimming instructions in the implant safety labeling.” 2)The consequences of not controlling, nor documenting, experimental parameters with variable impact on the results: “The scanner was allowed to make all of the RF adjustments automatically to avoid user induced variation.” 3)Stratifying the diverse experiments demonstrates equal temperature spread in a power vs B1+rms comparison: “We encourage the authors to view Fig. 4 and observe the vertical variation of RF power with RF coils for a given B1+rms.”
No takes yet. Share an insight, caveat, or question.
Steckner et al. (2026) studied this question. Stratifying diverse MRI experiments demonstrates equal temperature spread in a radiofrequency power versus B1+rms comparison near implanted conductive devices.
Synapse has enriched 10 related papers. Consider them for comparative context: