Why the study?
Do different THV designs and implantation strategies affect hydrodynamic performance during redo TAVI in bench testing?
Do different THV designs and implantation strategies affect hydrodynamic performance during redo TAVI in bench testing?
Bench testing demonstrates that redo TAVI generally yields favorable hydrodynamic performance, with optimal positioning depending on whether the failed valve's leaflets are intra- or supra-annular.
Answers are grounded in this paper and cite their sources.
Bench testing supports redo TAVI feasibility across designs; leaves open clinical translation and optimal positioning.
This study demonstrates that most THV designs and implantation strategies can result in favourable hydrodynamic performance following redo TAVI. Whether the leaflets of a failed THV are intra- or supra-annular may have important implications for the positioning of a redo-THV implant. Certain THV designs or implantation positions may be more desirable when performing redo TAVI.
No takes yet. Share an insight, caveat, or question.
Sathananthan et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: