Answers are grounded in this paper and cite their sources.
At 4° CP sampling, LT, MCSv, and LTMCS were found to be significantly correlated with VMAT dosimetric accuracy, expressed as Γ pass-rates. These parameters were found to be possible candidates for scoring plan complexity using threshold values. A finer CP separation (3°∕2°) led to a significant increase in dosimetric accuracy for plans with high leaf travel values, and to a decrease in correlation between LT and Γ passing rates. These results indicated that the influence of LT on VMAT dosimetric accuracy can be controlled by reducing CP separation. CP spacing for all plans requiring large leaf motion should not exceed 3°. The reported data were integrated to optimize our clinical workflow for plan creation, optimization, selection among rival plans, and patient-specific QA of VMAT treatments.
No takes yet. Share an insight, caveat, or question.
Masi et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: